UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM (2015-06-23

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
S/N: AY19258
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
California
UNITED STATES
A. UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEMB24802

Introduction

Affected Products

A. UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM

Reason

Beckman Coulter has confirmed low recovery values for Reticulocytes may be generated from the UniCel DxH 800 Coulter Cellular Analysis System due to intermittent high electrical noise on the Upper Median Angle Light Scatter (UMAL) produced by the volume, conductivity, scatter and multiple angle (VCSN) preamp board assembly part number (PN) b48187. The electrical noise produced by the VCSN preamp board assembly is greater than the instrument specification of 3.000 millivolt root mean square (MV RMS).