SIEMENS AXIOM ARTIS ZEE (2015-06-26) Jun 26, 2015 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number 10037 Manufacturer Siemens Ag Wittelsbacherplatz 2 Muenchen 80333 GERMANY H. AXIOM ARTIS FC 5904433 40033 Manufacturer Siemens Ag Wittelsbacherplatz 2 Muenchen 80333 GERMANY I. AXIOM ARTIS DFC MAGNETIC NAVIGATION 7727717 > 10 numbers, contact mfg Manufacturer Siemens Ag Wittelsbacherplatz 2 Muenchen 80333 GERMANY F. ARTIS ZEE CEILING 10094137 44102 Manufacturer Siemens Ag Wittelsbacherplatz 2 Muenchen 80333 GERMANY O. AXIOM ARTIS DBC 7728392 153146 Manufacturer Siemens Ag Wittelsbacherplatz 2 Muenchen 80333 GERMANY G. ARTIS ZEE BIPLANE 10094141 160190 Manufacturer Siemens Ag Wittelsbacherplatz 2 Muenchen 80333 GERMANY D.ARTIS ZEEGO MULTI-AXIS 10280959 > 10 numbers, contact mfg Manufacturer Siemens Ag Wittelsbacherplatz 2 Muenchen 80333 GERMANY E. ARTIS ZEE FLOOR 10094135 > 10 numbers, contact mfg Manufacturer Siemens Ag Wittelsbacherplatz 2 Muenchen 80333 GERMANY B. AXIOM ARTIS DTA 7008605 25006 Manufacturer Siemens Ag Wittelsbacherplatz 2 Muenchen 80333 GERMANY K. AXIOM ARTIS FA ALPHA 5904441 14026 Manufacturer Siemens Ag Wittelsbacherplatz 2 Muenchen 80333 GERMANY L. AXIOM ARTIS BC 5904649 49003 Manufacturer Siemens Ag Wittelsbacherplatz 2 Muenchen 80333 GERMANY C. AXIOM ARTIS DBC MAGNETIC NAVIGATION 5917054 28022 Manufacturer Siemens Ag Wittelsbacherplatz 2 Muenchen 80333 GERMANY M. AXIOM ARTIS BA 5904656 32070 Manufacturer Siemens Ag Wittelsbacherplatz 2 Muenchen 80333 GERMANY N. AXIOM ARTIS TA 7007755 53069 Manufacturer Siemens Ag Wittelsbacherplatz 2 Muenchen 80333 GERMANY A. AXIOM ARTIS DBA 7728392 35051 Manufacturer Siemens Ag Wittelsbacherplatz 2 Muenchen 80333 GERMANY J. AXIOM ARTIS DFC 7412807
Affected Products
AXIOM ARTIS DBA
AXIOM ARTIS DTA
AXIOM ARTIS DBC MAGNETIC NAVIGATION
ARTIS ZEEGO MULTI-AXIS
ARTIS ZEE FLOOR
ARTIS ZEE CEILING
ARTIS ZEE BIPLANE
AXIOM ARTIS FC
AXIOM ARTIS DFC MAGNETIC NAVIGATION
AXIOM ARTIS DFC
AXIOM ARTIS FA ALPHA
AXIOM ARTIS BC
AXIOM ARTIS BA
AXIOM ARTIS TA
AXIOM ARTIS DBC
If the tabletop is approximately fully extended at the head end and there is a large amount of liquid on the table surface, the liquid can make its way from the foot end of the tabletop onto the top cover of the table column where it can seep into the table through a gap. If this causes electronic components to become contaminated, table movements may no longer be possible.