ACUSON DIAGNOSTIC ULTRASOUND SYSTEMS (2015-06-30) Jun 30, 2015 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number >10, contact mfg Manufacturer Siemens Medical Solutions USA Inc. 685 East Middlefield Road California UNITED STATES B. ACUSON S3000 DIAGNOSTIC ULTRASOUND SYSTEM 10441730 >10, contact mfg Manufacturer Siemens Medical Solutions USA Inc. 685 East Middlefield Road California UNITED STATES A. ACUSON S2000 Diagnostic ultrasound system 10041461
Affected Products
A. ACUSON S2000 Diagnostic ultrasound system
B. ACUSON S3000 DIAGNOSTIC ULTRASOUND SYSTEM
Potential miscalculation when using the Doppler manual trace measurement tool in a full screen display format. The manual trace tool may incorrectly calculate the time average velocity (tav) and subsequently incorrectly calculate the pulsatility index (pi).