CLINITEK NOVUS ANALYZER (WITH RACK) (2015-07-02)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
S001046A. CLINITEK NOVUS ANALYZER (WITH RACK)
Manufacturer
Siemens Healthcare Diagnostics Manufacturing Ltd.,
Northern Road,
Sudbury, Suffolk, United Kingdom
CO10 2XQ
10494134

Introduction

Affected Products

A. CLINITEK NOVUS ANALYZER (WITH RACK) 

Reason

Siemens has identified a potential reporting error with the CLINITEK Novus Analyzer. If the report clarity setting is disabled, there is a potential for the specific gravity, Leukocytes, Albumin, p:c ratio, and/or the a:c ratio to be misreported. This issue impacts systems on SW Versions 1.1 and 1.2.