A. MICROSCAN MUELLER-HINTON BROTH WITH 3% LYSED HORSE BLOOD (LHB BROTH)
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
California
UNITED STATES
B1015-25
Introduction
Affected Products
A. MICROSCAN MUELLER-HINTON BROTH WITH 3% LYSED HORSE BLOOD (LHB BROTH)
Reason
Customers reported receiving LHB broth tubes which visually appeared to be discolored and/or turbid. Some of the tubes visually evaluated were confirmed to have been contaminated. The supplier performed retention samples testing for the two affected lots and found no contamination or turbid broth in any of the tubes; however Beckman Coulter is sending notification letters to customers who received affected lots for their awareness. If a contaminated tube were inadvertently used it could result in misinterpretation of susceptibility.