HEARTSTART XL+ (2015-06-25)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
>10, contact mfg.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
Massachusetts
UNITED STATES
A.HEARTSTART XL+861290

Introduction

Affected Products

A.HEARTSTART XL+

Reason

Through internal testing and customer complaint investigations, the following XL+ software and hardware issues have been identified:

Software:
  1. The XL+ may fail to complete the power on sequence and continuously reboot.
  2. The XL+ may either fail to power up or may shut down unexpectedly.
  3. The XL+ may have a software version that did not reset a fail-safe monitoring component which could delay of therapy or pacing interruption.
  4. The XL+ may fail to generate verbal prompts in AED mode.


Hardware:
  1. The XL+ may have been manufactured with a speaker that may fail.
  2. The battery may not seat properly causing the XL+ to shut down unexpectedly or remain powered on and not acknowledge or charge the battery.
  3. The XL+ exceeds the allowable radiated emissions level for class B Cispr11.
  4. The XL+ ECG signal from leads could be lost and unrecoverable.
  5. The XL+ Spo2 signal may lose communication and cause the device to reboot.
  6. The XL+ battery detection system may be disrupted and cause a false low battery alarm.