Health Products

GEMINI TF PET/CT IMAGING SYSTEM (2015-07-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
9224
Manufacturer
Philips Medical Systems (Cleveland) Inc.
595 MINER ROAD
CLEVELAND
Ohio
UNITED STATES
B. GEMINI TF BIG BORE PET/CT IMAGING SYSTEM4598 001 18561
7103
Manufacturer
Philips Medical Systems (Cleveland) Inc.
595 MINER ROAD
CLEVELAND
Ohio
UNITED STATES
A. GEMINI TF 64 PET/CT SYSTEM - MAIN UNIT4535 679 94741

Introduction

Affected Products
  1. GEMINI TF 64 PET/CT SYSTEM - MAIN UNIT
  2. GEMINI TF BIG BORE PET/CT IMAGING SYSTEM

Reason

Issue #1: When a single contour is saved in edit mode, TumorLoc does not save the contour in the correct location. The contour is saved in the location of the same slice where the last contour was drawn for the last region of interest (ROI) in the (ROI) list. A misplaced contour then results in a misplaced Isocenter or incorrect field size. This occurs when using the save button, exit button or send to laser button.

Issue #2: When the region of interest (ROI) is displayed on the Beam's Eye Viewer (BEV) in outline mode, and the (ROI) color is changed, the color is changed in the Multi-Planar Projection (MPR) viewers but not the (BEV) viewers.

Issue #3: TumorLoc created datasets are flipped and incorrectly labeled when non-gated data is loaded along with pulmonary gated data and is used to create intensity projection datasets (MIP, MINIP, or AVGIP). If the clinician obtains a clinical intensity projection dataset in a certain manner, a dataset can be generated that is flipped left to right. The resulting dataset is flipped and incorrectly labeled left-to-right when displayed in a viewer.

Issue #4: Duplicate region of interest (ROI) or Isocenter names can be created in TumorLoc. This duplication is only possible when using the copy feature or manually naming a (ROI) with the suffix "copy". This duplication can be created when creating a new Isocenter or (ROI), then creating another new Isocenter or (ROI) and naming the second Isocenter or (ROI) "first name copy" (either manually or by clicking copy on the first Isocenter or (ROI), where "first name" is the name of the first Isocenter or (ROI). If the user then selects the first Isocenter or (ROI) and clicks copy, the resulting copy has the same name as the second Isocenter/ROI: "first name copy".
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime