ADVIA CENTAUR CP SYSTEM ASSAY (2015-07-16) Jul 16, 2015 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type III
Lot or serial number Companies Brand name Model or catalog number ALL UNEXPIRED LOTS Manufacturer Siemens Healthcare Diagnostics Inc. 511 Benedict Ave Tarrytown New York UNITED STATES F.ADVIA CENTAUR SYSTEM - CALIBRATOR 04800646 38453 Manufacturer Siemens Healthcare Diagnostics Inc. 511 Benedict Ave Tarrytown New York UNITED STATES D.ADVIA CENTAUR SYSTEM - T3 ASSAY 04779663 ALL UNEXPIRED LOTS Manufacturer Siemens Healthcare Diagnostics Inc. 511 Benedict Ave Tarrytown New York UNITED STATES E.ADVIA CENTAUR SYSTEM - T4 ASSAY 110735 38453 Manufacturer Siemens Healthcare Diagnostics Inc. 511 Benedict Ave Tarrytown New York UNITED STATES B.ADVIA CENTAUR CP SYSTEM - T3 ASSAY 04779663 ALL UNEXPIRED LOTS Manufacturer Siemens Healthcare Diagnostics Inc. 511 Benedict Ave Tarrytown New York UNITED STATES C.ADVIA CENTAUR CADVIA CENTAUR SYSTEM -T4 ASSAY 110735 ALL UNEXPIRED LOTS Manufacturer Siemens Healthcare Diagnostics Inc. 511 Benedict Ave Tarrytown New York UNITED STATES A. ADVIA CENTAUR CP SYSTEM - CALIBRATOR A 04800646
Affected Products
A. ADVIA CENTAUR CP SYSTEM - CALIBRATOR A
B.ADVIA CENTAUR CP SYSTEM - T3 ASSAY
C.ADVIA CENTAUR CADVIA CENTAUR SYSTEM -T4 ASSAY
D.ADVIA CENTAUR SYSTEM - T3 ASSAY
E.ADVIA CENTAUR SYSTEM - T4 ASSAY
F.ADVIA CENTAUR SYSTEM - CALIBRATOR A
Siemens is issuing a customer notification for ADVIA centaur T4 assay for the manual 1:4 and 1:8 dilution recovery used on ADVIA Centaur and ADVIA Centaur CP Systems. Percent recoveries for 1:4 (with calibrator 'a' prior to lots ending in 87) and 1:8 (all lots) patient sample dilutions are lower than reported in the IFU. A similar shift between calibrator 'a' and the internal standardization for T3 assay was also identified.