Health Products

TECAN FREEDOM EVO ( 2015-05-31)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
1307002389
Manufacturer
Tecan Schweiz AG
Seestrasse 103
Mannedorf
8708
SWITZERLAND
B. FREEDOM EVO 10010641100
1303000035
Manufacturer
Tecan Schweiz AG
Seestrasse 103
Mannedorf
8708
SWITZERLAND
C. FREEDOM EVO 15010641150
1307003117
Manufacturer
Tecan Schweiz AG
Seestrasse 103
Mannedorf
8708
SWITZERLAND
A. FREEDOM EVO 20010641200

Introduction

Affected Products
  • A. FREEDOM EVO 200
  • B. FREEDOM EVO 100
  • C. FREEDOM EVO 150

Reason

Tecan detected a software bug that manifests itself when a specific set of conditions are met. This results in the door locks unlocking potentially allowing access for an operator to reach in while the arms are moving. This could potentially cause an injury to the operator if they are physically inside the freedom EVO’s working area during arm movement.