DIGITAL RADIOGRAPHY X-RAY SYSTEM DX-D 100 TYPE 5410 OR DX-D 100 WIRELESS TYPE 5411 (2015-06-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
> 10 numbers, contact mfg.
Manufacturer
AGFA Healthcare N.V.
Septestraat 27
Mortsel
2640
BELGIUM
A. DIGITAL RADIOGRAPHY X-RAY SYSTEM DX-D 100 TYPE 5410 OR DX-D 100 WIRELESS TYPE 5411SM-40HF-B-D-C

Introduction

Affected Products

A. DIGITAL RADIOGRAPHY X-RAY SYSTEM DX-D 100 TYPE 5410 OR DX-D 100 WIRELESS TYPE 5411 

Reason

When using the DX-D 100 unit, sporadic unintended movements caused by an electrostatic discharge of the unit to the ground may occur. Capacitive touch screens rely on the electrical properties of the human body to enable you to operate the device with very light touches of a finger. However, when liquid comes in contact with the touch screen it may incorrectly recognize this as user input.