Health Products

E CAM DUET SIGNATURE SERIES(2015-07-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
>10, contact mfg
Manufacturer
Siemens Medical Solutions USA Inc.
Molecular Imaging Group,
2501 North Barrington Road
Hoffman Estates
60192
Illinois
UNITED STATES
B. E CAM SIGNATURE SERIES7333250
>10, contact mfg
Manufacturer
Siemens Medical Solutions USA Inc.
Molecular Imaging Group,
2501 North Barrington Road
Hoffman Estates
60192
Illinois
UNITED STATES
A. E Cam DUET4381047

Introduction

Affected Products
  • A. E Cam DUET
  • B. E CAM SIGNATURE SERIES

Reason

A potential failure mode related to radial break and to perform an update to install a re-design of the brake redundancy circuit. The brake redundancy circuit is re-designed to force even a defective brake to engage in the unpowered state by applying a reverse charge to the brake and ensuring the brake has demagnetized.