Health Products

COBAS INTEGRA TINA-QUANT SOLUBLE TRANSFERRIN RECE (2015-07-03)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
60644901A. COBAS INTEGRA TINA-QUANT SOLUBLE TRANSFERRIN RECE
Manufacturer
Roche Diagnostics Gmbh
Sandhoferstrasse 116
Mannheim
68305
GERMANY
20763454122

Introduction

Affected Products

A. COBAS INTEGRA TINA-QUANT SOLUBLE TRANSFERRIN RECE

Reason

Roche received a customer complaint regarding elevated imprecision and too high recovery of STFR control set level 1 and DUP alarms during calibration with the reagent TQ Soluble Transferrin Receptor (STFR). The issue of elevated imprecision was confirmed by internal investigations on lot #606449 on Cobas C501 showed elevated imprecision of up to 8.9%. The other current reagent lots are not affected by this issue