A. FREESTYLE PRECISION NEO BLOOD GLUCOSE AND KETONE MONITORING SYSTEM
71385-75
Introduction
Affected Products
A. FREESTYLE PRECISION NEO BLOOD GLUCOSE AND KETONE MONITORING SYSTEM
Reason
Abbott Diabetes Care has become aware that the Freestyle Precision NEO Blood Glucose and Ketone Monitoring Systems may not function as intended. A meter firmware defect was identified affecting the basal titration (long-term insulin dose recommendation) feature for patients using long-acting basal insulin where the suggested insulin dose displays as 0U (units) instead of the basal insulin dose, if no titration occurs after the target timeframe (set by the HCP) ends and the average fasting blood glucose level is within the target range set by the HCP. Abbott Diabetes Care has determined that this issue does not pose a significant risk to users’ health or safety.