MICROSCAN LABPRO INFORMATION MANAGER SYSTEM (2015-07-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Version 1.0 to Version 4.11
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
California
UNITED STATES
A. MICROSCAN LABPRO INFORMATION MANAGER SYSTEM10361216

Introduction

Affected Products

A. MICROSCAN LABPRO INFORMATION MANAGER SYSTEM

Reason

Beckman coulter has confirmed the following issues can be encountered when generating organism identifications in the Microscan Labpro information manager system from dried overnight gram negative panels processed on Microscan Autoscan-4 Systems or read manually and entered into the Labpro System.

The use of dried overnight gram negative panels and the Labpro Software System on the Walkaway system is not affected. Some glucose non-fermenting organisms may display a color change (shade of yellow) in the GLU, SUC, and SOR at 48 hours. When this occurs:
  • Labpro software incorrectly allows the operator to manually edit the carbohydrate substrates (specifically GLU, SUC, and SOR) when manually reading dried overnight gram negative (conventional) panels with an ID hold status.
  • Labpro software incorrectly overwrites the previously stored carbohydrate substrate results (specifically GLU, SUC, and SOR) when reading dried gram negative overnight (conventional) panels which have an ID hold status on the Autoscan-4 System.
  • Labpro software then incorrectly uses the fermenter identification substrates to calculate the biotype which may cause non-fermenters such as an Acinetobacter species to be incorrectly identified as fermenters, for example, a Shigella species.