DIMENSION CREATININE SYSTEMS (2015-07-21) Jul 21, 2015 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type III
Lot or serial number Brand name Companies Model or catalog number >10 NUMBERS, CONTACT MFR D. DIMENSION EXL SYSTEM CREATININE (CREA) Manufacturer Siemens Healthcare Diagnostics Inc. 500 Gbc Drive, Mailstop 514 Newark, Delaware 19714-6101 UNITED STATES DF33A >10 NUMBERS, CONTACT MFR B. DIMENSION SYSTEM CREATININE (CREA) Manufacturer Siemens Healthcare Diagnostics Inc. 500 Gbc Drive, Mailstop 514 Newark, Delaware 19714-6101 UNITED STATES DF33A >10 NUMBERS, CONTACT MFR C. DIMENSION XPAND SYSTEM CREATININE (CREA) Manufacturer Siemens Healthcare Diagnostics Inc. 500 Gbc Drive, Mailstop 514 Newark, Delaware 19714-6101 UNITED STATES DF33A >10 NUMBERS, CONTACT MFR A. DIMENSION VISTA SYSTEM CREATININE (CREA) Manufacturer Siemens Healthcare Diagnostics Inc. 500 Gbc Drive, Mailstop 514 Newark, Delaware 19714-6101 UNITED STATES K1033
Affected Products
A. DIMENSION VISTA SYSTEM CREATININE (CREA)
B. DIMENSION SYSTEM CREATININE (CREA)
C. DIMENSION XPAND SYSTEM CREATININE (CREA)
D. DIMENSION EXL SYSTEM CREATININE (CREA)
Siemens has confirmed hemoglobin (HB) interference (greater than 10 percent) for dimension Crea above 200 mg/dl HB at a creatinine concentration of 1.5 mg/dl and for dimension vista Crea above 200 mg/dl HB at a creatinine concentration of 1.5 mg/dl. Current claims for HB interference are less than 10 percent at these Crea and HB concentrations.