ICHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEMS (2015-07-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
> 10 numbers, contact mfg.B. ICHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEM W/LOAD/UNLOAD
Manufacturer
Iris Diagnostics, a Division of Iris International Inc.
9172 Eton Avenue
Chatsworth
UNITED STATES
800-7167
V02564A. ICHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEM
Manufacturer
Iris Diagnostics, a Division of Iris International Inc.
9172 Eton Avenue
Chatsworth
UNITED STATES
800-7719

Introduction

Affected Products
  1. ICHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEM
  2. ICHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEM W/LOAD/UNLOAD

Reason

IRIS International has confirmed the iChemVELOCITY Automated Urine Chemistry System does not maintain the on-board strip stability of 5 days at 18-28 degrees c at 20-80% relative humidity (RH) as stated in the iChemVELOCITY product labeling. The on-board strip stability is affected by temperature and humidity in the laboratory. If a laboratory operates outside of the parameters defined in table 1 of the urgent medical device recall letter, test results may be affected as indicated below:
  • false low or false negative results: ketone, blood
  • false high or false positive results: bilirubin, glucose, leukocytes
  • false high results: urobilinogen


ascorbic acid, protein, Ph, and nitrite are not affected.