Health Products

NDI DISPOSABLE REFLECTIVE MARKER SPHERE (2015-07-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
C100641501
Manufacturer
Northern Digital Inc
103 Randall Drive
Waterloo
N2V1C5
Ontario
CANADA
A. NDI DISPOSABLE REFLECTIVE MARKER SPHERE41773

Introduction

Affected Products

A. NDI DISPOSABLE REFLECTIVE MARKER SPHERE

Reason

On April 30, 2015, NDI received a complaint from a customer reporting that a DRMS sphere fell apart while the patient's leg was being aligned with the navigation system during a total knee replacement surgery. Based on the reported complaint; NDI promptly instituted a stop ship order, and then performed a recall of the affected lot. Further investigation has identified a similar failure mode, with different root causes that have occurred at a significantly lower rate in additional lots of the product.