Carescape R860 (Evair component) (2015-07-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
>10, contact mfg.
Manufacturer
Datex-Ohmeda
Inc.
3030 Ohmeda Drive
Wisconsin
UNITED STATES
A. Carescape R860 (Evair component)1506-8600-000

Introduction

Affected Products

A. Carescape R860 (Evair component)

Reason

The following issue is being addressed by GE healthcare file no. FMI 34066: the EVair component manufacturer, Imtmedical AG, is conducting a field corrective action. Imtmedical has identified that potential friction of cables and tubing internal to the EVair compressor may result in an inability for the compressor to output compressed air. There is a potential for temporary oversupply of oxygen to the patient as a result of this failure because the GE Healthcare ventilators are designed to ventilate with 100% oxygen in the case of loss of input air supply pressure.