HEMODIALYSIS CATHETER SETS (2015-06-23) Jun 23, 2015 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number All Manufacturer Medical Components Inc, Also trading as Medcomp 1499 Delp Drive Harleysville 19438 Pennsylvania UNITED STATES D. SILICON CATHETER SETS MC101241 All Manufacturer Medical Components Inc, Also trading as Medcomp 1499 Delp Drive Harleysville 19438 Pennsylvania UNITED STATES B. HEMO-CATH SILICONE DOUBLE LUMEN TEMPORARY ACCESS CATHETER SETS SL15E All Manufacturer Medical Components Inc, Also trading as Medcomp 1499 Delp Drive Harleysville 19438 Pennsylvania UNITED STATES C. PEDIATRIC HEMO-CATH SILICONE DOUBLE LUMEN TEMPORARY ACCESS SL12P All Manufacturer Medical Components Inc, Also trading as Medcomp 1499 Delp Drive Harleysville 19438 Pennsylvania UNITED STATES A. HEMO-CATH SILICONE DOUBLE LUMEN TEMPORARY ACCESS CATHETERS SL15C
Affected Products
HEMO-CATH SILICONE DOUBLE LUMEN TEMPORARY ACCESS CATHETERS HEMO-CATH SILICONE DOUBLE LUMEN TEMPORARY ACCESS CATHETER SETS PEDIATRIC HEMO-CATH SILICONE DOUBLE LUMEN TEMPORARY ACCESS SILICON CATHETER SETS
Cracks in silicon extension tubing of hemodialysis catheters propagate over time resulting in holes because of inadequate use during the insertion procedure, that is, the extension line clamp is clamped onto the guide wire. The hole in the extension tubing leaks when the device is being pressurized with a syringe during flushing or during dialysis treatment.