Health Products

ALLSET GOLD HIGH RES SSP UNITRAY (2015-07-08)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
012 1467113B. A HIGH RES SSP UNITRAY 10TESTS
Manufacturer
One Lambda Inc
21001 Kittridge Street
Canoga Park
California
UNITED STATES
47290-10
012 1395079A. ALLSET GOLD A HIGH RES SSP
Manufacturer
One Lambda Inc
21001 Kittridge Street
Canoga Park
California
UNITED STATES
54010D

Introduction

Affected Products
  1. ALLSET GOLD A HIGH RES SSP
  2. A HIGH RES SSP UNITRAY 10TESTS

Reason

A customer reported that in allset+ gold a high res SSP kit, lot 012 1608823 the reactivity for lane 86 was negative for an a*03:20 allele when the labeling indicates the reactivity should be positive. Also, the High Res SSP UniTray kits are affected by this complaint since they contain the same reagents as the allset+ gold a high res SSP kits. The labeling discrepancy may lead to potentially mistyping a*03:20 allele (population frequency < 0.01%) as a*03:01.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime