Health Products

BUCKY DIAGNOST (2015-07-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
394898/SN08000161
Manufacturer
Philips Medical Systems Dmc Gmbh
Rontgenstrabe 24
Hamburg
22335
GERMANY
A. BUCKY DIAGNOST9890 010 83821

Introduction

Affected Products

A. BUCKY DIAGNOST

Reason

The trauma supporting block (connecting the trauma arm with the ceiling suspension telescope) is fixed with 3 screws, which are also used to adjust the correct alignment of it. Philips has become aware that in case this is not done in the correct way, the trauma arm is not fastened securely anymore and could fall down.