BRILLIANCE AND INGENUITY CT SYSTEMS (2015-07-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
30003B. INGENUITY CT SYSTEM
Manufacturer
Philips Medical Systems (Cleveland) Inc.
595 MINER ROAD
CLEVELAND
Ohio
UNITED STATES
728323
9920A. BRILLIANCE 40/64 SLICE CT SYSTEM
Manufacturer
Philips Medical Systems (Cleveland) Inc.
595 MINER ROAD
CLEVELAND
Ohio
UNITED STATES
4550 110 04011

Introduction

Affected Products
  1. BRILLIANCE 40/64 SLICE CT SYSTEM
  2. INGENUITY CT SYSTEM

Reason

Some scans unexpectedly set the axial (Z axis) scan length to a greater length than the user set in the plan. The issue occurs only when the following 3 conditions are met:-the plan employs dual-surview, the plan requests either a sagittal or a coronal reconstruction, the gantry is tilted