Health Products

SOFLEX UV-ABSORBING SILICONE POSTERIOR CHAMBER IO (2015-06-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
> 1000 contact Manufacturer
Manufacturer
Bausch and Lomb Inc.
1400 NORTH GOODMAN STREET
Rochester
14609
New York
UNITED STATES
A. SOFLEX UV-ABSORBING SILICONE POSTERIOR CHAMBER IOLI61AO

Introduction

Affected Products

A. SOFLEX UV-ABSORBING SILICONE POSTERIOR CHAMBER IO

Reason

Based on a limited number of complaints received for broken Haptics during lens loading and insertion, an investigation was initiated.  This investigation resulted in Bausch & Lomb determining that a portion of a lot of Haptic material is performing differently than other lots in term of material elongation properties.