Health Products

I-Portal Neuro Otologic Test Center (NOTC)(2015-04-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
SN#900152
Manufacturer
Aim Instrumentation ltd
5232 IRMIN STREET
Burnaby
V5J 1Y7
CANADA
A. I-Portal Neuro Otologic Test Center (NOTC)I-PORTAL NOTC-P

Introduction

Affected Products

A. I-Portal Neuro Otologic Test Center (NOTC)

Reason

Neuro Kinetics Inc. (NKI) has issued a voluntary device field correction related to a spontaneous unexpected chair rotation on one unit in the field. The unexpected chair rotation may happen upon unit start up. If the end user ignores software error message and continues with pressing "seating" or "testing" buttons. The error message is displayed before any involvement of a patient.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime