ASP200S/ASP300S TISSUE PROCESSOR(2015-08-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
5541
Manufacturer
Leica Microsystems Nussloch GMBH
HEIDELBERGER STREET 17-19
Nussloch
69222
GERMANY
A. ASP200S/ASP300S TISSUE PROCESSORASP200S

Introduction

Affected Products

A. ASP200S/ASP300S TISSUE PROCESSOR

Reason

One complaint was received where the remote alarm of an affected instrument did not alert the customer properly of an error that occurred during processing. A review meeting was held by the manufacturer to determine that the drawing of the remote alarm cable does have an error which led to a malfunction of the remote alarm. Cables with this error were used since April 24, 2013 in production. A ship hold of the ASP200s/ASP300s was initiated on July 13, 2015. Within this time period 901 instruments were distributed globally with the affected cables. The affected devices have an incorrect internal wiring to the remote alarm plug, which can cause the remote alarm to not function correctly in the case of an instrument failure.