Health Products

(VEPTR) VERTICAL EXPANDABLE PROSTHETIC TITANIUM RIB CAUDAL RIB SUPPORTS (2015-08-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
>100 NUMBERS, CONTACT MFG
Manufacturer
Synthes USA LLC
1101 SYNTHES AVENUE
Monument
80132
Colorado
UNITED STATES
A. (VEPTR) VERTICAL EXPANDABLE PROSTHETIC TITANIUM RIB CAUDAL RIB SUPPORTS497.087

Introduction

Affected Products

A. (VEPTR) VERTICAL EXPANDABLE PROSTHETIC TITANIUM RIB CAUDAL RIB SUPPORTS

Reason

This recall is being initiated because the last hole on the VEPTR Caudal Rib Supports, subject to this recall, may be partially formed. Fully lengthening the construct to the partially formed last hole may result in mechanical failure of the construct due to extension bar breakage, which could require unplanned surgical intervention to exchange components. In addition to the risks inherent to re-operation, potential effects on the patient could also include pain, loss of deformity correction, and/or soft tissue/visceral injury.