Alaris Syringe Model 8110 (2015-07-09)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
contact manufacturer
Manufacturer
Carefusion 303, Inc.
10020 Pacific Mesa Blvd.
San Diego
92121-2733
California
UNITED STATES
A. Alaris Syringe Model 81108110

Introduction

Affected Products

A. Alaris Syringe Model 8110

Reason

After error code 351.6740 and channel error is cleared on the PCU, the syringe pump is unresponsive to key presses until the next power cycle, or the module is detached and re-attached.