One complaint was received from a customer in Spain where poor processing was identified and samples were deemed non-diagnosable. the investigation of the unit when returned to the manufacturer on July 15th led to a safety meeting which determined that the instrument is incorrectly getting into a state of "bottle empty" during retort filling, after passing the fill lever sensor 2 and prior to reaching level sensor 3, thus not completing a 5 litre fill.