LEICA ASP6025 TISSUE PROCESSOR(2015-08-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
314
Manufacturer
Leica Microsystems Nussloch GMBH
HEIDELBERGER STREET 17-19
Nussloch
69222
GERMANY
A. LEICA ASP6025 TISSUE PROCESSORASP6025

Introduction

Affected Products

A. LEICA ASP6025 TISSUE PROCESSOR

Reason

One complaint was received from a customer in Spain where poor processing was identified and samples were deemed non-diagnosable. the investigation of the unit when returned to the manufacturer on July 15th led to a safety meeting which determined that the instrument is incorrectly getting into a state of "bottle empty" during retort filling, after passing the fill lever sensor 2 and prior to reaching level sensor 3, thus not completing a 5 litre fill.