Health Products

BD MAX SYSTEM SOFTWARE V4.44A(2015-07-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
N/A
Manufacturer
Becton Dickinson and Company
7 Loveton Circle
Sparks
Maryland
UNITED STATES
A. BD MAX SYSTEM SOFTWARE V4.44A441916

Introduction

Affected Products

A. BD MAX SYSTEM SOFTWARE V4.44A

Reason

BD Max System Software version 4.44a contains a software anomaly that has the potential to affect some open system reagent (OSR) customers utilizing user defined protocol (UDP) assays. Customers utilizing only IVD assays are not affected. This anomaly may cause OSR customer system to incorrectly switch the columns of the truth table on the user's display. If the user tries to fix their display data, the display will appear correct however, the database may become incorrect, potentially resulting in incorrect results reporting.