Health Products

CLOSED SUCTION SYSTEMS (2015-08-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
M5082T614
Manufacturer
Kimberly-Clark Global Sales LLC
1400, Holcomb Bridge Road
Roswell
ÉTATS-UNIS 30076
Georgia
UNITED STATES
B. KIMVENT MULTI-ACCESS PORT CLOSED SUCTION SYSTEM F22103-5
M5082T502
Manufacturer
Kimberly-Clark Global Sales LLC
1400, Holcomb Bridge Road
Roswell
ÉTATS-UNIS 30076
Georgia
UNITED STATES
A. KIMVENT NEONATAL / PEDIATRIC "Y" TRACHEAL SUCTION1910

Introduction

Affected Products

A. KIMVENT NEONATAL / PEDIATRIC "Y" TRACHEAL SUCTION

B. KIMVENT MULTI-ACCESS PORT CLOSED SUCTION SYSTEM F

Reason

Halyard Health has recently received reports indicating that the thumb valve of a very small percentage of Kimvent Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "locked" position