Health Products

LX3 MICROSCOPE FLOOR STAND used with Luxor Microscope (2015-08-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
S/N: 1501088003X1
Manufacturer
Alcon Laboratories Inc.
6201 South Freeway
Fort Worth
Texas
UNITED STATES
A. LX3 MICROSCOPE FLOOR STAND used with Luxor Microscope8065752560

Introduction

Affected Products
  1. LX3 MICROSCOPE FLOOR STAND used with Luxor Microscope

Reason

Alcon is conducting a voluntary medical device correction of all LX3 microscope floor stands after receiving reports concerning loss of functionality due to issues with the cabling between the power, host and illumination modules, potentially causing the system to display a system message, and/or resulting in illumination failure. Alcon will retrofit the cable design on any impacted lx3 microscope floor stands.