Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
0101B. IMMULITE 1000 SYSTEM - C-PEPTIDE ASSAY
Manufacturer
Siemens Healthcare Diagnostics Products Limited
Glyn Rhonwy
Llanberis
LL55 4EL
UNITED KINGDOM
LPEPCVM
0101A. IMMULITE SYSTEM C-PEPTIDE ASSAY
Manufacturer
Siemens Healthcare Diagnostics Products Limited
Glyn Rhonwy
Llanberis
LL55 4EL
UNITED KINGDOM
LPEPCVM

Introduction

Affected Products
  1. IMMULITE SYSTEM C-PEPTIDE ASSAY
  2. IMMULITE 1000 SYSTEM - C-PEPTIDE ASSAY

Reason

Manufacturer has confirmed the potential for over-recovery of the c-peptide CVM top level with CVM lots 012 (level j) and 0101 (level 4) when used with the Immulite Systems C-Peptide Assays. in some instances this may lead to results outside the range published in the CVM instructions for use (IFU). Internal testing demonstrated that all other levels for CVM lots 012 and 0101 are unaffected. The potential over-recovery is limited to the top level CVM and patient results are not impacted.