Health Products

BECKMAN COULTER AU IMMUNOCHEMISTRY SYSTEM ANALYZERS (2015-07-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
>10 numbers, contact mfg.
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
California
UNITED STATES
B. BECKMAN COULTER AU IMMUNOCHEMISTRY SYSTEM - AU480 ANALYZERAU480
>10 numbers, contact mfg.
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
California
UNITED STATES
C. BECKMAN COULTER AU5800 CLINICAL CHEMISTRY ANALYZERAU5810
>10 numbers, contact mfg.
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
California
UNITED STATES
A. BECKMAN COULTER AU IMMUNOCHEMISTRY SYSTEM - AU680 INSTRUMENTAU680

Introduction

Affected Products

A. BECKMAN COULTER AU IMMUNOCHEMISTRY SYSTEM - AU680 INSTRUMENT

B. BECKMAN COULTER AU IMMUNOCHEMISTRY SYSTEM - AU480 ANALYZER

C. BECKMAN COULTER AU5800 CLINICAL CHEMISTRY ANALYZER

Reason

In March 2011, the material and molding process of the coolant cup (C cups) was changed. Since the change, Beckman Coulter has received complaints regarding leaking and cracked C cups. When this occurs, coolant can leak onto and damage one of the printed circuit boards (PCBS) in the analyzers.