B. BECKMAN COULTER AU IMMUNOCHEMISTRY SYSTEM - AU480 ANALYZER
AU480
>10 numbers, contact mfg.
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
California
UNITED STATES
C. BECKMAN COULTER AU5800 CLINICAL CHEMISTRY ANALYZER
AU5810
>10 numbers, contact mfg.
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
California
UNITED STATES
A. BECKMAN COULTER AU IMMUNOCHEMISTRY SYSTEM - AU680 INSTRUMENT
AU680
Introduction
Affected Products
A. BECKMAN COULTER AU IMMUNOCHEMISTRY SYSTEM - AU680 INSTRUMENT
B. BECKMAN COULTER AU IMMUNOCHEMISTRY SYSTEM - AU480 ANALYZER
C. BECKMAN COULTER AU5800 CLINICAL CHEMISTRY ANALYZER
Reason
In March 2011, the material and molding process of the coolant cup (C cups) was changed. Since the change, Beckman Coulter has received complaints regarding leaking and cracked C cups. When this occurs, coolant can leak onto and damage one of the printed circuit boards (PCBS) in the analyzers.