Health Products

ACUSON SC2000 ULTRASOUND SYSTEM - MAINFRAME (2015-08-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
400319, 401223
Manufacturer
Siemens Medical Solutions USA
Inc.
685 East Middlefield Road
California
UNITED STATES
A. ACUSON SC2000 ULTRASOUND SYSTEM - MAINFRAME10433816

Introduction

Affected Products

A. ACUSON SC2000 ULTRASOUND SYSTEM - MAINFRAME

Reason

The Acuson SC2000 ultrasound system considers uppercase/lowercase differences in the same patient name as unique patient instances when registered on the same ultrasound system. The system recalls the case sensitivity even if the original study has been removed from the hard drive. if these differences are not corrected at the time of registration, the system does not capture images or clips.
 