RE-FLEX ROTATE AND ADAPTERS (2015-07-10) Jul 10, 2015 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Brand name Companies Model or catalog number N/A F. Flex-Foot Assure Manufacturer Ossur HF Grjothals 5 Reykjavik 110 ICELAND >10, CONTACT MFG N/A G. Balance Foot J Manufacturer Ossur HF Grjothals 5 Reykjavik 110 ICELAND >10, CONTACT MFG N/A D. Vari-Flex XC Manufacturer Ossur HF Grjothals 5 Reykjavik 110 ICELAND >10, CONTACT MFG N/A E. Re-Flex Shock Manufacturer Ossur HF Grjothals 5 Reykjavik 110 ICELAND >10, CONTACT MFG N/A B. Talux Manufacturer Ossur HF Grjothals 5 Reykjavik 110 ICELAND >10, CONTACT MFG N/A C. Vari-Flex Manufacturer Ossur HF Grjothals 5 Reykjavik 110 ICELAND >10, CONTACT MFG N/A A. Re-Flex Rotate Manufacturer Ossur HF Grjothals 5 Reykjavik 110 ICELAND >30, CONTACT MFG
Affected Products
A. Re-Flex Rotate
B. Talux
C. Vari-Flex
D. Vari-Flex XC
E. Re-Flex Shock
F. Flex-Foot Assure
G. Balance Foot J
Recent findings suggest that non-ossur female mating adapters connected to ossur ankle male pyramids have caused early failure of the male pyramid connector. This is due to incompatible geometry of the female mating adaptors used and/or improper set up of the prosthesis.