A. 3M DuraPrep Surgical Solution Patient Preoperative Skin Preparation
8630 8635
Introduction
Affected Products
A. 3M DuraPrep Surgical Solution Patient Preoperative Skin Preparation
Reason
During investigation into a complaint of discoloured Duraprep solution, 3m determined that the product solution had unintended exposure to the sterilant during package sterilization. This type of exposure would only be possible if the glass ampule holding the solution is compromised with a crack. Direct exposure to the sterilant causes a color change from dark brown to yellow/clear and renders the product unfit for use given formation of by-products with the potential for sensitization, irritation and acute dermal toxicity.