B. UNICEL DXI 600 ACCESS IMMUNOASSAY SYSTEM - FREE T3 ASSAY
A13422
431433
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
California
UNITED STATES
C. UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM - FREE T3 ASSAY
A13422
431433
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
California
UNITED STATES
A. ACCESS SYSTEM - FREE T3 ASSAY
A13422
Introduction
Affected Products
A. ACCESS SYSTEM - FREE T3 ASSAY
B. UNICEL DXI 600 ACCESS IMMUNOASSAY SYSTEM - FREE T3 ASSAY
C. UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM - FREE T3 ASSAY
Reason
Beckman Coulter has determined that some Access Free T3 Reagent Packs (P/N A13422) of lot 431433 were filled incorrectly. These packs contained insufficient quantity of reagents in one or more of the pack wells. Affected reagent packs have the potential to generate incorrect control and patient results dependent upon the instrument and software version installed at the time the reagent lot was in use.