C. DISCOVERY MR750W 3.0T MAGNETIC RESONANCE IMAGING
DISCOVERY MR750W 3.0T
00000279069MR9
Manufacturer
Ge Medical Systems LLC
3200 North Grandview Blvd
Waukesha
UNITED STATES
A. 3.0T SIGNA HDX MR SYSTEM - MAIN UNIT
M3335JJ
Introduction
Recalled Products
3.0T SIGNA HDX MR SYSTEM - MAIN UNIT
DISCOVERY MR750 MR SYSTEM - MAIN UNIT
DISCOVERY MR750W 3.0T MAGNETIC RESONANCE IMAGING
Reason
The predicted Head SAR value from the modeling, when using GE Healthcare 3.0t MR scanners with the whole body RF transmit coil and receive-only surface coils for head and/or neck imaging, suggests that actual SAR delivered to the head could potentially exceed the 3.2 W/kg SAR limit defined in iec60601-2-33. This issue does not affect MR scans performed outside the head and neck areas.