Introduction

Affected Products
  1. Pressure Monitoring Set
  2. Pressure Monitoring Kit with VAMP
  3. Pressure Monitoring Set

Reason

Through post market surveillance, flow restriction was detected from all arterial, CVP and PA lines during priming and flushing. Flow restriction was found inside IV trifurcated adaptor. The IV tubing lumen was blocked by the adaptor and restricted flow paths to extension tubes and pressure lines.