Health Products

MONACO (2015-09-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Elekta Business Area Software Systems Impac Medical Systems
Inc.
13723 Riverport Drive
Missouri
UNITED STATES
MONACOMONACO

Introduction

Affected Products

MONACO

Reason

Two issues affecting the Monaco device: 1) when CT images are created this latter way and imported in Monaco, the image is displayed correctly but the electron densities are incorrectly flipped in both the right-left and superior-inferior directions. 2) when using multiple prescription and the first prescription has forced densities, these forced densities are not being maintained with the second prescription.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime