Elekta Business Area Software Systems Impac Medical Systems
Inc.
13723 Riverport Drive
Missouri
UNITED STATES
MONACO
MONACO
Introduction
Affected Products
MONACO
Reason
Two issues affecting the Monaco device: 1) when CT images are created this latter way and imported in Monaco, the image is displayed correctly but the electron densities are incorrectly flipped in both the right-left and superior-inferior directions. 2) when using multiple prescription and the first prescription has forced densities, these forced densities are not being maintained with the second prescription.