Health Products

VANCOMYCIN ASSAYS (2015-09-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lots.B. VANC2 (ONLINE TDM VANCOMYCIN FOR ROCHE/HITACHI COBAS C 311 AND C 501/502 SYSTEMS
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
4491050190
All lots.C. VANCOMYCIN (ONLINE TDM VANCOMYCIN FOR ROCHE/HITACHI 912 AND MODULAR P ANALYZERS)
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
4642481190
All lots.A. VANC2 (ONLINE TDM VANCOMYCIN FOR ROCHE/HITACHI COBAS C 311 AND C 701/702 SYSTEMS
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
5842310190

Introduction

Affected Products
  1. VANC2 (ONLINE TDM VANCOMYCIN FOR ROCHE/HITACHI COBAS C 311 AND C 701/702 SYSTEMS
  2. VANC2 (ONLINE TDM VANCOMYCIN FOR ROCHE/HITACHI COBAS C 311 AND C 501/502 SYSTEMS
  3. VANCOMYCIN (ONLINE TDM VANCOMYCIN FOR ROCHE/HITACHI 912 AND MODULAR P ANALYZERS)

Reason

An inaccurately low result was reported for a Vancomycin patient sample on Cobas C501. This was confirmed by Roche. Detailed investigations revealed that the sample contains a component which inhibits the drug labeled enzyme glucose-6-phosphate dehydrogenase (Vanc-G6PDH) in the emit assay.