SYNCHRON SYSTEMS (2015-09-08)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Beckman Coulter, Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
B. Synchron Systems Enzymatic Creatinine Reagent (CR-E)A60298
All lots.
Manufacturer
Beckman Coulter, Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
A. Uric Acid Reagent (URIC)442785

Introduction

Affected Products
  1. Uric Acid Reagent (URIC)
  2. Synchron Systems Enzymatic Creatinine Reagent (CR-E)

Reason

N-Acetyl Cysteine (NAC) has been identified as an interference source for certain assays involving Trinder Chemical Pathways. Beckman Coulter initiated an investigation to determine if the interference would be observed in the DxC Synchron products that utilize trinder reaction. The results from the study showed that CR-E and uric reagents are significantly affected by therapeutic levels of N-Acetyl Cysteine. NAC is currently not listed as an interferent in the chemistry information sheets for CR-E or for uric, which may lead to false low results for CR-E and uric to be generated.