AXIOM ARTIS MODULAR ANGIOGRAPHY SYSTEMS (2015-09-14) Sep 14, 2015 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type III
Lot or serial number Companies Brand name Model or catalog number Not applicable Manufacturer Siemens AG Wittelsbacherplatz 2 Muenchen 80333 GERMANY H. AXIOM ARTIS MP DIAGNOSTIC 5904466 Not applicable Manufacturer Siemens AG Wittelsbacherplatz 2 Muenchen 80333 GERMANY I. AXIOM ARTIS BC 5904649 Not applicable Manufacturer Siemens AG Wittelsbacherplatz 2 Muenchen 80333 GERMANY F. AXIOM ARTIS DFC 7412807 Not applicable Manufacturer Siemens AG Wittelsbacherplatz 2 Muenchen 80333 GERMANY G. AXIOM ARTIS FA ALPHA 5904441 Not applicable Manufacturer Siemens AG Wittelsbacherplatz 2 Muenchen 80333 GERMANY D. AXIOM ARTIS FC 59 04 433 Not applicable Manufacturer Siemens AG Wittelsbacherplatz 2 Muenchen 80333 GERMANY E. AXIOM ARTIS DFC MAGNETIC NAVIGATION 7727717 Not applicable Manufacturer Siemens AG Wittelsbacherplatz 2 Muenchen 80333 GERMANY B. AXIOM ARTIS DMP 7555365 Not applicable Manufacturer Siemens AG Wittelsbacherplatz 2 Muenchen 80333 GERMANY K. AXIOM ARTIS DBC 7728392 Not applicable Manufacturer Siemens AG Wittelsbacherplatz 2 Muenchen 80333 GERMANY C. AXIOM ARTIS DBC MAGNETIC NAVIGATION 59 17 054 Not applicable Manufacturer Siemens AG Wittelsbacherplatz 2 Muenchen 80333 GERMANY A. AXIOM ARTIS DBA 7555357 Not applicable Manufacturer Siemens AG Wittelsbacherplatz 2 Muenchen 80333 GERMANY J. AXIOM ARTIS BA 5904656
Affected Products
AXIOM ARTIS DBA
AXIOM ARTIS DMP
AXIOM ARTIS DBC MAGNETIC NAVIGATION
AXIOM ARTIS FC
AXIOM ARTIS DFC MAGNETIC NAVIGATION
AXIOM ARTIS DFC
AXIOM ARTIS FA ALPHA
AXIOM ARTIS MP DIAGNOSTIC
AXIOM ARTIS BC
AXIOM ARTIS BA
AXIOM ARTIS DBC
It was identified that there is potential that a possible position sensor fault in the swivel base axis is not being detected by the system software. Following the enabling of movement by the operator, the C-Arm System can unexpectedly move faster than normal. Please note that the system will not move without movement being initiated by the operator, however, the speed at which it can move would be faster than normal.