AXIOM ARTIS MODULAR ANGIOGRAPHY SYSTEMS (2015-09-14)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
H. AXIOM ARTIS MP DIAGNOSTIC5904466
Not applicable
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
I. AXIOM ARTIS BC5904649
Not applicable
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
F. AXIOM ARTIS DFC7412807
Not applicable
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
G. AXIOM ARTIS FA ALPHA5904441
Not applicable
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
D. AXIOM ARTIS FC59 04 433
Not applicable
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
E. AXIOM ARTIS DFC MAGNETIC NAVIGATION7727717
Not applicable
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
B. AXIOM ARTIS DMP7555365
Not applicable
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
K. AXIOM ARTIS DBC7728392
Not applicable
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
C. AXIOM ARTIS DBC MAGNETIC NAVIGATION59 17 054
Not applicable
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
A. AXIOM ARTIS DBA7555357
Not applicable
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
J. AXIOM ARTIS BA5904656

Introduction

Affected Products
  1. AXIOM ARTIS DBA
  2. AXIOM ARTIS DMP
  3. AXIOM ARTIS DBC MAGNETIC NAVIGATION
  4. AXIOM ARTIS FC
  5. AXIOM ARTIS DFC MAGNETIC NAVIGATION
  6. AXIOM ARTIS DFC
  7. AXIOM ARTIS FA ALPHA
  8. AXIOM ARTIS MP DIAGNOSTIC
  9. AXIOM ARTIS BC
  10. AXIOM ARTIS BA
  11. AXIOM ARTIS DBC

Reason

It was identified that there is potential that a possible position sensor fault in the swivel base axis is not being detected by the system software. Following the enabling of movement by the operator, the C-Arm System can unexpectedly move faster than normal. Please note that the system will not move without movement being initiated by the operator, however, the speed at which it can move would be faster than normal.