Health Products

WELCH ALLYN SURESIGHT VISION SCREENER(2015-08-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
200700393WELCH ALLYN SURESIGHT VISION SCREENER
Manufacturer
Welch Allyn Inc.
4341 State Street Road
Skaneateles Falls
13153-0220
New York
UNITED STATES
14016

Introduction

Affected Products

WELCH ALLYN SURESIGHT VISION SCREENER

Reason

During service, SureSight Vision Screeners were calibrated and the calibration activity performed may not have returned the device to working order. In its current state, the device will most likely not be able to take a reading or it may be very difficult to get a reading. Devices affected are limited to those serviced between July 17, 2015 and August 7, 2015 at the Welch Allyn Skaneateles Falls location.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime