Health Products

MRXPERION MR INJECTION SYSTEM(2015-08-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Serial number 100029
Manufacturer
Bayer Medical Care Inc.
1 Bayer Drive
Indianola
15051-0780
Pennsylvania
UNITED STATES
MRXPERION MR INJECTION SYSTEMMRXP 200

Introduction

Affected Products

MRXPERION MR INJECTION SYSTEM

Reason

It was identified that during an injection there is a potential that a saline phase could not be completed, if the injection is started or resumed while a keep vein open (KVO) pulse is active. Additionally, a software revision will be implemented to adjust the reminder behavior to ensure that if the injection phase is paused manually, the reminder timer resumes once the injection phase is resumed.