ALLURA XPER UNIT(2015-09-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
F. ALLURA XPER FD20 OR TABLE000454
All lots.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
D. ALLURA XPER FD10ALLURA XPER FD10
All lots.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
E. ALLURA XPER FD20 SYSTEM - MAIN UNITALLURA XPER FD20
All lots.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
B. ALLURA XPER FD20/10 SYSTEM - MAIN UNITALLURA XPER FD20/10
All lots.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
C. ALLURA XPER FD20/20ALLURA XPER FD20/20
All lots.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6
Best
5684
NETHERLANDS
A. ALLURA XPER FD10/10 - MAIN UNITALLURA XPER FD10/10

Introduction

Affected Products
  1. ALLURA XPER FD10/10 - MAIN UNIT
  2. ALLURA XPER FD20/10 SYSTEM - MAIN UNIT
  3. ALLURA XPER FD20/20
  4. ALLURA XPER FD10
  5. ALLURA XPER FD20 SYSTEM - MAIN UNIT
  6. ALLURA XPER FD20 OR TABLE

Reason

Philips has become aware that memory problems on the Embedded Technology eXtended (ETX®)-module in the system interface board (SIB) box cause intermittent failures. The SIB box plays a role in starting and stopping and controlling the image functionality. Although highly intermittently, the failure can occur both during start up as well as during system use. The failure does not affect an active X-ray run and the run can be finished. However, no new run can be started until the system is rebooted. As such, the failure has no impact on the critical moment in a critical procedure.