Dash and Tram 451 Patient Monitors and Patient Data Module (2013-02-26) Feb 26, 2013 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type III
Lot or serial number Brand name Companies Model or catalog number More than 100 numbers, contact manufacturer. D. Patient Data Module Manufacturer GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee Wisconsin UNITED STATES PDM-XXX-XXX-XXXX More than 100 numbers, contact manufacturer. E. Moniteur patient Dash 4000 Manufacturer GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee Wisconsin UNITED STATES DASH4-XXXX-XXXX-XXXX More than 100 numbers, contact manufacturer. B. Dash 5000 Patient Monitor Manufacturer GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee Wisconsin UNITED STATES DASH5-XXXX-XXXX-XXXX More than 100 numbers, contact manufacturer. C. Tram 451 Patient Monitor Manufacturer GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee Wisconsin UNITED STATES T451=A More than 100 numbers, contact manufacturer. A. Dash 3000 Patient Monitor Manufacturer GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee Wisconsin UNITED STATES DASH3-XXXX-XXXX-XXXX
Affected Products
A. Dash 3000 Patient Monitor
B. Dash 5000 Patient Monitor
C. Patient Data Module
D. Patient Data Module
E. Dash 4000 Patient Monitor
ECG and respiration signal noise caused by Dash, PDM and/or Tram ECG cable connector movement may result in reduced ECG, respiration and arrhythmia detection performance.