Health Products

RadPRO 40kW - Digital Mobile X-Ray System (2015-07-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
S/N between G29393 and G62480
Manufacturer
Virtual Imaging, Inc.
720 SOUTH POWERLINE ROAD, SUITE E
DEERFIELD BEACH
33442
Florida
UNITED STATES
RadPRO 40kW - Digital Mobile X-Ray SystemMOBILE 3 40KW 50G/C

Introduction

Affected Products

RadPRO 40kW - Digital Mobile X-Ray System

Reason

Electrostatic energy may be stored in systems that can cause a short-circuit which, in turn, can generate an electromagnetic pulse which might disturb the electronics of the digital motion control board producing uncontrolled movements and such uncontrolled movements could cause the system to bump into the user or patient.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime