Introduction

Affected Products

STERRAD NX STERILIZATION SYSTEM

Reason

Advanced Sterilization Products (ASP) has recently determined the current STERRAD NX System is exceeding the electromagnetic compatibility class radiated emissions specification limit by 20 decibel micro-volt per meter, which equates to 0.001 volt per meter at a distance of 10 meters. The risk is assessed as low for interference to active implantable and wearable devices and other nearby electrical equipment.