Plum 360 Infusion System with MedNet (2015-10-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Various serial number contact manufacturer.Plum 360 Infusion System with MedNet
Manufacturer
Hospira Inc.
275 North Field Drive
Lake Forest
60045
Illinois
UNITED STATES
30010-13

Introduction

Affected Products

Reason

When the user selects "yes" at the "new patient?" screen, and the program was not changed prior to powering off the infuser, the distal occlusion pressure setting is assigned a value of 15 psi (775 mmHg) instead of the default value. In addition, the user interface will incorrectly display the default distal occlusion pressure value. The default value is set at 6 psi from the manufacturer. The user would not be immediately aware of the unintended occlusion pressure setting unless the user views the current distal pressure value on the infuser display and it is greater than the distal pressure alarm limit. This may result in delay of therapy related to delay in identifying the occlusion cause.